ReFlex Power™ Programming Manual AC Module Remote Programming
92 M380056-03 Rev L
REPLY
DESCRIPTION
Unable to change the Voltage Range -
Hardware Prohibited, or
User tries to change the voltage range
with
SOUR<n>:VOLT:RANGE <bool>
Unable to change the Voltage Range
Output Enabled or Aux Relay Closed
Unable to open Module Data file - file
system problem, or
User tries to store the module data
information and calibration files with
CAL<n>:STORE
5 POINT CAL FILE WRITING ERROR,
or
CAL FILE WRITING ERROR
4.4.2 E
RROR
/E
VENT
Q
UEUE
The Error/Event Queue is a FIFO (First-In-First-Out) buffer for storing detected
errors. When the queue is read with the SYSTem:ERRor? query, the error that
was detected first will be read first, and the error that was detected last will be
read last. When all errors are read, the message 0, “No error” will be returned.
Up to 10 errors can be stored before overflow would occur. Overflow will result in
error code -350, “Queue overflow” overwriting the last entry. All subsequent
errors will be ignored until the queue is read, and space is again available.