Ametek ReFlex Programming Manual Manuel d'utilisateur Page 88

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 134
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 87
ReFlex Power™ Programming Manual DC Module Remote Programming
84 M380056-03 Rev M
The Tracking modes do NOT function in RAMP functions, and faults may be generated. Tracking
modes must be turned off and OVP, OCP, and UVP set to proper values for the specific RAMP
command, else faults may occur. After the RAMP command has completed, and new setpoints
issued, Tracking may be then turned back on.
2
The average List/SCPI command that is compiled into a List executable takes from 1 to 4 instruction
index units, and complex SCPI commands such as STATe 1, takes 16 instruction index units,
because of relay, module power on, and setpoint sequencing. A voltage ramp takes 3 instruction
index units. An OUTP:STAT 0 command takes 7 units.
Vue de la page 87
1 2 ... 83 84 85 86 87 88 89 90 91 92 93 ... 133 134

Commentaires sur ces manuels

Pas de commentaire