DC Module Remote Programming ReFlex Power™ Programming Manual
M380056-03 Rev L 65
3.11
STATUS SCPI COMMANDS
3.11.1
STATUS
SCPI
C
OMMAND
S
UMMARY
STATus[n]
:MODE
:DELAY <0+NR1>
:DELAY?
:MODule
:ENABles <0+NR1>
:ENAB?
:FAULts?
3.11.2
STATUS
SCPI
C
OMMAND
R
EFERENCE
STATus<n>
Status subsystem. Module ‘n’.
:MODE:DELAY <0+NR1>
Sets the delay timer for responding to a status change from CV
to CC mode. Value in counts: 0 to 255, of approximately 20ms
for each count. Delay counter is immediately reloaded when the
mode switches back to CV.
Default timer value is 0.
:MODE:DELAY?
Returns the CV to CC status delay timer value. *RST sets to 0.
:MODule
:ENABles<0+NR1>
Enable supervisory. See Table 3-3 for enable values.
:ENABles?
Query the enabled supervisories
:FAULts?
Query the module faults
Table 3-3. DC Fault and Enable Register
Bit
Hex Value
Description
0 0x00000001
LDF Line Drop Fault
1 0x00000002
RSF Reverse Sense Fault
2 0x00000004
OCP Over Current Fault
3 0x00000008
OVP Over Voltage Protection Fault
4 0x00000010
ROCP Redundant Over Current Protection Fault
5 0x00000020
ROVP Redundant Over Voltage Protection Fault
6 0x00000040
UVP Under Voltage Protection Fault
7 0x00000080
RESERVED
8 0x00000100
IMPTEMPHSK1 Heatsink1 Temperature is Warm
9 0x00000200
IMPTEMPHSK2 Heatsink2 Temperature is Warm
10 0x00000400
OVTEMPHSK1 Heatsink1 Temperature Exceeds Limit
11 0x00000800
OVTEMPHSK2 Heatsink2 Temperature Exceeds Limit
12 0x00001000
AFF Air Flow Fault
13 0x00002000
UVP5V Internal 5V Under Voltage Fault
14 0x00004000
OVP5V 5V Over Voltage Fault