ReFlex Power™ Programming Manual DC Module Remote Programming
58 M380056-03 Rev M
:LOCK
:STORe
:UNLock <string>
:REVERT:FACTory
3.6.2 CALIBRATION
SCPI
C
OMMAND
R
EFERENCE
Command
Description
CALibrate<n>
Calibration subsystem. Module ‘n’.
:INITial
:CURRent
[:AMPLitude] <NRf>
Sets the power-on default value of current.
[:AMPLitude]?
Query the power-on default value of current
:PROTection <NRf>
Sets the power-on default value of the over current
protection.
:PROTection?
Query the power-on default value of the over current
protection.
:VOLTage
[:AMPLitude] <NRf>
Sets the power-on default voltage. Should be set to zero
volts to avoid hot switching of output relay.
[:AMPLitude]?
Query the power-on default voltage. Should be set to zero
volts to avoid hot switching of output relay.
:PROTection <NRf>
Sets the power-on default value of the over voltage
protection.
:PROTection?
Query the power-on default value of the over voltage
protection.
:STATe <bool>
1
Set the initial POR output state of the unit
:STATe?
Query the initial POR output state setting of the unit
:UNDERVOLTage
:PROTection <NRf>
Sets the power-on default value of the under voltage
protection.
:PROTection?
Query the power-on default value of the under voltage
protection.
:DEFAult
Sets initial settings to Factory defaults of 0.0
Volts/Amps/UVP, OVP/OCP to Module Max values and
POR State to Off.
:OUTPut
:CURRent
:COUNTS <NRi>
Set Current DAC
setpoint, in counts
:FIVEPOINT<1-5> <0+NRf>
Sets the current output calibration point (1 through 5). The
actual output current is measured with an external meter.
:FIVEPOINT?
Returns 5 pairs of current calibration setpoint count and
related current values
:PROTection
:COUNTS <NRi>
Set Protection value 0-4095
:VOLTage
:COUNTS <NRi>
Set Voltage DAC setpoint, in counts
:FIVEPOINT<1-5> <NRf>
Sets the voltage output calibration point (1 through 5). The
actual output voltage is measured with an external meter.