AC Module Remote Programming ReFlex Power™ Programming Manual
M380056-03 Rev L 103
Command
Description
:RANGe?
Query the voltage range of the unit.
:PROTection<NRf>
Sets the Over -Voltage Protection setpoint.
:PROTection?
Returns the Over Voltage Protection setpoint.
:UPSET
4
<NRf>,<NRf>,
<NRf>,<NRf>
The UPSET Event has four arguments that allow surge/sag or dropouts to be
programmed to start at a specific phase delay, to a specified RMS voltage
level, for specified time, and then apply an after event voltage level.
:UPSET?
4
Returns the above UPSET Event parameters if the event has not finished.
The returned duration value will be zero when the event is complete.
:UNDERVOLTage
:PROTection[<NRf>]
Sets the Under Voltage Protection setpoint.
:PROTection?
Returns the Under Voltage Protection setpoint.
:TRIPped?
Query whether Under Voltage Protection supervisory tripped or not
:STATe <bool>
Sets or clears the global detect enable for the Under Voltage Protection
supervisory. 0 = disable, 1 = enable
:STATe?
Query the global detect enable for the Under Voltage Protection supervisory.
0 = disabled, 1 = enabled
:MODE
:SHUTdown<bool>
Set or clear the shutdown enable for Under Voltage Protection supervisory
:SHUTdown?
Query the shutdown enable for Under Voltage Protection supervisory
1
See Section 4.12 for examples for using Group operation.
2
PHASE:LEAD specifies that the specified PHASE <NRf> angle value is the number of
degrees that the rising waveform zero cross will occur before the Phase Reference in time
.
Cleared by *RST, by command, or Power cycle. Power on default is LAG.
2
PHASE:LAG specifies that the specified PHASE <NRf> angle value is the number of
degrees that the rising waveform zero cross will occur after the Phase Reference in time.
Cleared by *RST, by command, or Power cycle. Power on default is LAG.
The Phase Reference is a timing signal that is generated by the Phase Master Module. It is
used by the Phase Master and Slave Modules to offset their own rising zero cross waveform
timing from the Phase Reference time marker. The PHASE setpoint accepts values
between -360.0 and +360.0 degrees, this will interact with LEAD/LAG.
3
Prior to setting the Voltage above 125V in the low range or above 250V in the high range, you
must set the three Current related setpoints, SOURce<n>:CURRent, SOURce<n>:CURRent
:PROTection and SOURce<n>:INRushcurr:PROTection, else a “parameter out of range”
error will occur. The reason is that these three Current setpoints are designed to limit the
maximum continuous power to 875VA, and have the protection values at safe levels; therefore,
they must be lowered in order to raise the voltage above the values stated for the low range or
for the high range. The three Current setpoint formulae in order, are as follows:
Step 1: Current Setpoint (I
set
) as a function of voltage:
Low Range formula: I
set
(v) = I
set
(max) * 125 / V
set
, for V
set
> 125
Example: to set the voltage to 137V. I
set
(137) = 7.14 * 125/137 = 6.514A(max)
SOUR<n>:CURR 6.51
Vue de la page 106
1 2 ... 102 103 104 105 106 107 108 109 110 111 112 ... 133 134

Commentaires sur ces manuels

Pas de commentaire

InLine 16431Q manuels

Manuels d'utilisation et guides de l'utilisateur pour Extensions de puissance InLine 16431Q.
Nous fournissons des manuels en pdf 1 InLine 16431Q à télécharger gratuitement par type de document : Fiche technique


Inline 16431Q Fiche technique (1 pages)


marque: Inline | Catégorie: Power extensions | Taille: 0.06 MB |

 

Table des matières





Plus de produits et de manuels pour Extensions de puissance InLine

Modèles Type de document
16481T Fiche technique   InLine 16481T power extension, 1 pages
16481W Fiche technique   InLine 16481W power extension, 1 pages
16461Q Fiche technique   InLine 16461Q power extension, 1 pages
16485T Fiche technique   InLine 16485T power extension, 1 pages
16441 Fiche technique   InLine 16441 power extension, 1 pages
16465T Fiche technique   InLine 16465T power extension, 1 pages
16445T Fiche technique   InLine 16445T power extension, 1 pages
16441T Fiche technique   InLine 16441T power extension, 1 pages
16435S Fiche technique   InLine 16435S power extension, 1 pages
16463V Fiche technique   InLine 16463V power extension, 1 pages
16485W Fiche technique   InLine 16485W power extension, 1 pages
16443T Fiche technique   InLine 16443T power extension, 1 pages
16485S Fiche technique   InLine 16485S power extension, 1 pages
16463S Fiche technique   InLine 16463S power extension, 1 pages
16441W Fiche technique   InLine 16441W power extension, 1 pages
16463T Fiche technique   InLine 16463T power extension, 1 pages
16445W Fiche technique   InLine 16445W power extension, 1 pages
16461I Fiche technique   InLine 16461I power extension, 1 pages
16445S Fiche technique   InLine 16445S power extension, 1 pages
16461V Fiche technique   InLine 16461V power extension, 1 pages