SG Series Programming IEEE/RS232 Configuration and Remote Programming
M550129-03 Rev K 3-3
3.2
RS232 SETUP PROCEDURE
This section provides a quick reference for the configuration requirements for
RS232. Refer to Sections 3.3.1 and 3.3.2 for detailed information on the rear
panel configuration switches.
Parameter Setting Notes
Baud Rate SGA models: Fixed, 19200 The baud rate for SGI models is
selectable through the front panel
using the Remote Menu; refer to
Section 3.8.11 of the SGI Operation
Manual, M550221-01.
SGI models: Selectable,
2400 to 19200
Data Bits 8
Stop Bits 1
Parity None
Flow Control None
Incoming Termination
Character
CR (Carriage Return):
HEX, 0x0d (DEC, 13)
Outgoing Termination
Character(s)
CR LF (Carriage Return and
Line Feed):
HEX, 0x0d 0x0a (DEC, 13 10)
Selectable using SCPI command,
SYST:NET:TERM; refer to
Section 6.9.2.
1. Build an RS232 communications cable per the pinout description
illustrated in Figure 3-4 (with crossover of signals Rx/Tx and CTS/RTS):
RJ-type 6P6C Plug
Terminal
D-Subminiature 9-Pin Connector
Female Socket
1 2
2 7
3 No Connection
4 5
5 3
6 8
Figure 3-4. RS232 Communications Cable Pinout