Ametek SFA Series Ethernet Programming Manuel d'utilisateur Page 123

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 159
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 122
SG Series Programming SCPI Command Operation
M550129-03 Rev K
6-51
EXAMPLE 4
Issue *RST to establish a known state.
In this example, create a sequence that pauses at its end to let the last settings remain in
effect. To do this use the PAUSE command.
Now create a new SEQ3:
PROG:SEL:NAME “SEQ3”
PROG:MALL DEFAULT
PROG:DEF 1, VIMODE, 3, 4, 11, 10 //go to 3 volts, 4 amps, 11 volts ovp, for
10 sec
PROG:DEF 2, RAMPTOV, 3, 5, 4, 11, 10 //ramp from 3 volts to 5 volts in 10 sec
PROG:DEF 3, VIMODE, 5, 4, 11, 10 //hold 5 volts for 10 sec
PROG:DEF 4, RAMPTOV, 5, 3, 4, 11, 10 //ramp from 5 volts to 3 volts in 10 sec
PROG:DEF 5, VIMODE, 3, 4, 11, 10 //hold 3 volts for 10 sec
PROG:DEF 6, SUBCALL, SEQ2 //call SEQ2 as a subsequence
PROG:DEF 7, VIMODE, 4, 5, 11, 6 //go to 4 volts, 5 amps, 11 volts ovp, for 6 sec
PROG:DEF 8, PAUSE //allow the output to remain at the last setting
PROG:DEF 9, STOP //stop running the sequence while the
unit remains at the state of the last
command within the sequence.
PROG:STAT COMPLETE
PROG:SAVE:SEL
PROG:STAT?
Response: Ram[44]=“STOPPED”,Slave[44]=”STOPPED”