SG Series Programming SCPI Command Operation
M550129-03 Rev K
6-57
EXAMPLE 6
This example shows how to make an infinite loop for the creation of a continuous square
wave:
Issue *RST to establish a known state.
PROG:NAME Square Wave
PROG:MALL DEFAULT
PROG:DEF 1, VIMODE, 0, 5, 15, 0.5 //go to 0 volts, 5 amps, 15 volts ovp, for
0.5 seconds
PROG:DEF 2, VIMODE, 10,5,15,0.5 //go to 10 volts, 5 amps, 15 volts ovp, for
0.5 seconds
PROG:DEF 3,GOTO,Square Wave //go to top of this sequence and loop
indefinitely)
PROG:STAT COMPLETE
PROG:SAVE:SEL
PROG:STAT?
Response: Ram[110]=“STOPPED”,Slave[110]=”STOPPED”
OUTP:STAT ON
PROG:STAT RUN
The output for Example 6 should look something like the following:
Figure 6-6. Power Supply Output for Example 6
etc…