Calibration SG Series Programming
7-10
M550129-03 Rev K
9. Reset the output current to 0 amps.
CAL:OUTP:CURR:DAC 0
10. Program the Digital Interface Adapter (DIA) to calculate the current readback
calibration gain and offset values:
CAL:OUTP:CURR:CALC
CAL:MEAS:CURR:CALC
11. Program the DIA to unlock the non-volatile memory for calibration value
storage:
CAL:UNLOCK “6867”
12. Program the Digital Interface Adapter (DIA) to store the date:
GPIB:
CAL:DATE YYYYMMDD
Ethernet:
CAL:MOD:LASTCALDATE MM DD YYYY
CAL:MOD:NEXTCALDATE MM DD YYYY
13. Program the DIA to store the calibration values in non-volatile memory:
CAL:STORE
Allow 15 seconds for the non-volatile memory to be updated with the
new calibration values.
14. Program the DIA to lock the non-volatile memory for calibration value
protection
CAL:LOCK
Calibration is complete.