:PAUSe Set harmonics pause.
:TABle Select Individual harmonic table 1,2,3 or ALL.
:IHARmonics
:LEVel Set interharmonics level.
:DWELl Set interharmonics dwell.
:PAUSe Set interharmonics pause.
IEC413:STATe
This command will enable the IEC413 and initialize the associated data. Note that user
defined test levels for the USER group can only be programmed when the IEC413 state is
OFF.
Command Syntax IEC413:STATe <state>
Parameters <ON|OFF|1|0 >
Examples IEC413:RUN STAT ON
Query Syntax IEC413:STAT?
Returned Parameters <boolean>
IEC413:RUN
This command will run the test. The parameters will define the section of test to be run.
Command Syntax IEC413:RUN <group>
Parameters < ALL|GROup|RANGe|POINt >
Examples IEC413:RUN ALL
Query Syntax IEC413:RUN?
Returned Parameters <group>
IEC413:PAUSe
This command will cause the test to pause at the time the command is issued.
Command Syntax IEC413:PAUSe <boolean>
Parameters ON|OFF|1|0
Examples IEC413:PAUS ON
Query Syntax IEC413:PAUS?
Returned Parameters <boolean>
IEC413:CLASs
This command will select the class. There are two predefined classes 2 and 3. Class 1 is
supported through the use of the USER defined class. Other tests levels can be set in the
USER class as well.
Command Syntax IEC413:CLASs <class>
Parameters 2|3|USER
Examples IEC413:CLAS 3
Query Syntax IEC413:CLAS?
Returned Parameters <class>