AMETEK Programmable Power BPS / MX / RS Series SCPI Programming Manual
108 Manual P/N 7003-961 Rev AA
VOLTage
Phase Selectable
This command programs the AC rms, DC or offset output voltage level of the power source.
The voltage mode must be set for AC to accept an AC voltage. The Voltage mode must also
be set to AC+DC to accept a DC offset value.
Command Syntax [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude][:AC] <NRf+>
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]:DC] <NRf+>
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]:OFFSet<NRf
+>
Parameters 0 to maximum voltage range specified by the LIMit:VOLTage
command
Unit V (rms voltage)
*RST Value 0 volt
Examples VOLT 250 VOLT:LEV 25
Query Syntax [SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Returned Parameters <NR2>
Related Commands [SOUR:]MODE VOLT:TRIG
VOLTage:TRIGgered
Phase Selectable
This command selects the AC rms or DC amplitude that the output voltage will be set to
during a triggered step or pulse transient.
Command Syntax SOURce:]VOLTage[:LEVel]:TRIGgered:AMPLitude]<NRf+>
Parameters 0 to maximum voltage range specified by the LIMit:VOLTage
command
Unit V (rms voltage)
*RST Value 0 volt
Examples VOLT:TRIG 120 VOLT:LEV:TRIG 120
Query Syntax SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
Returned Parameters <NR2>
Related Commands VOLT VOLT:MODE