BPS / MX / RS Series SCPI Programming Manual AMETEK Programmable Power
Manual P/N 7003-961 Rev. AA 103
PONSetup:VOLTage:MODE
This command sets the initial voltage mode at the power on. This mode is DC or AC
Command Syntax [SOURce:]PONSetup: VOLTage:MODE
Parameters AC | DC | ACDC [ACDC on 3Pi Controller Only]
Examples PONS: VOLTage:MODE DC
Query Syntax PONS: VOLTage:MODE?
Returned Parameters <CRD>
PONSetup:VRANge
This command sets the initial voltage range of the power source.
Command Syntax [SOURce:]VRANge<NRf+>
Parameters 150200 | 300 | 400
Examples PONS:VRAN 150
Query Syntax PONS:VRAN?
Returned Parameters <NR2>
PONSetup:WGRoup [3Pi Controller Only]
This command determines the source of its waveform storage. It has a total of 4 groups.
Change of group requires power down, an IEEE-488 Device Clear (not available when usign
the RS232C interface) or a *RST command to take effect.
Command Syntax [SOURce:]PONSetup:WGRoup <n>
Parameters 0|1|2|3
Examples PONSetup:WGRoup 2
Query Syntax PONSetup:WGR?
Returned Parameters <NR1>
Related Commands TRACe:CAT?