AMETEK Programmable Power BPS / MX / RS Series SCPI Programming Manual
120 Manual P/N 7003-961 Rev AA
SYSTem:CONFigure:NOUTput
Sets the mode configuration of the power source to single-phase or a three-phase mode.
The MODE option must be enabled for this command to work. If MODE option is not
available, the query response will be FIX.
Command Syntax SYSTem:CONFigure:NOUTput <noutput mode>
Parameters ONEPhase | THReephase
Example SYST:CONF:NOUT ONE
Query Syntax SYSTem:CONF?
Query response ONE, THREE or FIX Note: Query response CRD differs
from parameter format on this command.
Returned Paramters <CRD>
SYSTem:ETIMe?
This command will return the total number of accumulated hours, minutes and seconds.
Command Syntax SYSTem:ETIMe?
Parameters none
Example SYST:ETIM?
Returned Parameters <NR1>,<NR1>,<NR1>
SYSTem:TEMPerature?
This command will return the internal ambient temperature of the AC/DC power source in
degrees Celsius.
Command Syntax SYSTem:TEMPerature?
Parameters none
Example SYST:TEMP?
Returned Parameters <NR2>
SYSTem:COMMunicate:GPIB:ADDRess
This command will set the GPIB address.
Command Syntax SYSTem:COMMunicate:GPIB:ADDRess <NRF>
Parameters 0 to 31 <listen address>
Example SYST:COMM:GPIB:ADDR 2
Returned Parameters <NR1>
SYSTem:COMMunicate:LAN:ADDRess
This command will set the LAN IP address. The query format returns the IP address set. To
query the actual IP address used by the LAN interface (in case the IP address is not static
but assigned by a DHCP server), the NCON parameter (Network Connection Setting) must
be added to the query. The default value is 0.0.0.0.