Ametek SFA Series Programming Manual Manuel d'utilisateur Page 39

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 48
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 38
Sorensen SF Series IEEE 488.2 and SCPI Command Operation
6. Program the output of the second calibration point to approximately 85% of full scale
voltage by sending the following command string from the computer:
CAL:OUTP:CURR:DAC 55700
7. Let the output settle and measure the voltage with the meter.
8. Enter the actual voltage readback corresponding to the DAC value 5570 of the second
calibration point:
CAL:MEAS:VOLT:POINT 2 <voltage>
9. Reset the output voltage to 0 volts by programming the current to 0 amps.
CAL:OUTP:CURR:DAC 0
10. Program the DIA to calculate the voltage readback calibration gain and offset values:
CAL:MEAS:VOLT:CALC
11. Program the DIA to unlock the non-volatile memory for calibration value storage:
CAL:UNLOCK “6867”
12. Program the DIA to store the calibration values in non-volatile memory:
CAL:STORE
13. Program the DIA to lock the non-volatile memory for calibration value protection
CAL:LOCK
14. The voltage readback calibration is complete. The unit may be turned OFF or other
calibration procedures may be performed.
4.4 Current Programming Calibration
STEP
DESCRIPTION
1. Program the overcurrent protection to maximum to prevent nuisance trips:
CAL:OUTP:CURR:PROT:DAC 65535
3. Program the output of the first calibration point to approximately 15% of full scale current
by sending the following command string from the computer:
CAL:OUTP:CURR:DAC 9830
M550292-03 (Programming) 4-3
Vue de la page 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 47 48

Commentaires sur ces manuels

Pas de commentaire