Elgar CW-P Series SCPI Commands
Programming Manual 3-3
:FREQuency <value>
Set the frequency of the CW.
:FREQuency?
Returns the frequency setpoint of the CW.
:PHASe <value>
Adjust phase 1, 2, or 3 phase angle offset from
external trigger.
:PHASe?
Return the present phase angle setpoint.
:VOLTage?
Return the present voltage setpoint.
:VOLTage [<value>]
Set the voltage setpoint.
:LEVel?
Return the present voltage setpoint.
:LEVel [<value>]
Set the voltage setpoint.
:IMMediate?
Return the present voltage setpoint immediately
without waiting for any other commands to be
processed.
:IMMediate [<value>]
Set the voltage setpoint immediately without waiting
for any other commands to be processed.
:AMPLitude <value>
Set the voltage setpoint.
:AMPLitude?
Return the present voltage setpoint.
:PROTection?
Return the overvolt protection of the CW.
Always returns „1‟.
:PROTection [<value>]
Set the overvolt protection setpoint value.
:LEVel <value>
Set the overvolt protection setpoint value.
:LEVel?
Return the overvolt protection value.
:STATe <boolean>
Specify the overvolt protection state. 1 = protection
turned on. 0 will be ignored.
:STATe?
Return the present overvolt state of the CW. Always
will return „1‟.
:TRIPped?
Return the present state of the overvolt protection
circuit. 0 = there has been no overvolt condition,
1 = CW has shut down due to an overvolt condition.
:RANGe <value>
Set the voltage range to HIGH range or LOW range
:RANGe?
Returns the present set voltage range of the CW.
Return values will be either „1‟ or „0‟.
:RMSLOOP <boolean>
If set to „1‟ turns on RMS loop software.
:RMSLOOP?
Returns „1‟ if RMS loop software is enabled.