SLM-Series AC/DC
Electronic Load
Operating and Programming
Manual
Vue de la page 0
1 2 3 4 5 6 ... 86 87

Résumé du contenu

Page 1 - Electronic Load

Xantrex Technology Inc.Programmable Products9250 Brown Deer RoadSan Diego, CA 92121USA858 450 0085 Phone858 458 0257 Faxwww.programmablepower.comSmart

Page 2

This page intentionally left blank. WA-2

Page 3

v Important Safety Instructions This manual has been written expressly for the Sorensen brand SLM series of electronic loads, which have been designed

Page 4

vi This page intentionally left blank.

Page 5

v CONTENTS SECTION 1 FEATURES, FUNCTIONS, AND SPECIFICATIONS ...1-1 1.1 General Description ...

Page 6

Contents SLM-Series AC/DC Load vi 3.2.2 Frequency Setting: ...3-6 3.2.3

Page 7 - SAFETY SYMBOLS

SLM-Series AC/DC Load Contents viiFigure 2-2 Connection Method for Small Load Current Condition...2-4 Figure 2-3 Typica

Page 8

Contents SLM-Series AC/DC Load viii This page intentionally left blank.

Page 9

M540072-01 Rev B 1-1 SECTION 1 FEATURES, FUNCTIONS, AND SPECIFICATIONS 1.1 GENERAL DESCRIPTION SLM-Series, AC/DC Electronic Load is used to test the

Page 10

Features, Functions, Specifications SLM-Series AC/DC Load 1-2 M540072-01 Rev B 125V100V Figure 1-2 SLM-150-8-300 Electronic Load Power Curve 250V200

Page 11

SLM-Series AC/DC Load Features, Functions, and Specifications M540072-01 Rev B 1-3 Figure 1-4 SLM-500-1-300 Electronic Load Power Curve

Page 13 - CONTENTS

Features, Functions, Specifications SLM-Series AC/DC Load 1-4 M540072-01 Rev B 1.1.2 Operating Modes The work mode of SLM-Series AC/DC Electronic Lo

Page 14 - Load GPIB/RS-232 Operating

SLM-Series AC/DC Load Features, Functions, and Specifications M540072-01 Rev B 1-5 1.2 FEATURES AND FUNCTIONS • Modular design facilitates fast an

Page 15

Features, Functions, Specifications SLM-Series AC/DC Load 1-6 M540072-01 Rev B 1.4 SPECIFICATIONS NOTE: The following specifications apply 25°±5°:

Page 16

M540072-01 Rev B 2-1 SECTION 2 INSTALLATION AND MAINTENANCE 2.1 INTRODUCTION This section discusses the installation and removal procedures for the

Page 17 - AND SPECIFICATIONS

Installation SLM-Series AC/DC Load 2-2 M540072-01 Rev B Figure 2-1 Plug-in Installation and Removal 2.2.2 Removal 1. Turn the SLM mainframe power

Page 18

SLM-Series AC/DC Load Installation M540072-01 Rev B 2-3 2.3 CONNECTIONS CAUTION: PREVENT DAMAGE TO THE LOAD. Do NOT apply voltage or current wit

Page 19

Installation SLM-Series AC/DC Load 2-4 M540072-01 Rev B 2.3.2 Voltage Sensing Input (Vsense) BNC Connector. To solve the conductor voltage drop und

Page 20 - 1.1.2 Operating Modes

M540072-01 Rev B 3-1 SECTION 3 OPERATION 3.1 FRONT PANEL CONTROLS AND INDICATORS This section describes the front panel and its manual operation of

Page 21 - 1.3 ACCESSORIES

Operation SLM-Series AC/DC Electronic Load 3-2 M540072-01 Rev B 1 Model Label Shows load model series and its voltage, current and power specifica

Page 22 - 1.5 REGULATORY COMPLIANCE

SLM-Series AC/DC Electronic Load Operation M540072-01 Rev B 3-3 • During FREQ ON condition (see 15): o For frequency (“FrEq”) setting, displays D

Page 24 - 2.2.4 Service or Repair

Operation SLM-Series AC/DC Electronic Load 3-4 M540072-01 Rev B o VA meter with the unit as “VA”.. • Lower DM displays the lower limit of: o Vo

Page 25 - 2.3 CONNECTIONS

SLM-Series AC/DC Electronic Load Operation M540072-01 Rev B 3-5 17 Vsense BNC, Voltage Sensing Input BNC Connector. To solve the conductor voltage

Page 26 - 2.4 MAINTENANCE

Operation SLM-Series AC/DC Electronic Load 3-6 M540072-01 Rev B Power: Press the LIMIT key until the LIMIT LED is on and the “W” LED (5) is lit.

Page 27 - OPERATION

SLM-Series AC/DC Electronic Load Operation M540072-01 Rev B 3-7 Zero CrossingSYNCInputSYNC "OFF"+5V330ΩSYNCEXT+EXT-SYNC "ON" F

Page 28

Operation SLM-Series AC/DC Electronic Load 3-8 M540072-01 Rev B 5. Press the “FREQ” key two times to exit setting mode. 6. Press the appropriate

Page 29

SLM-Series AC/DC Electronic Load Operation M540072-01 Rev B 3-9 SetG/NG VALimitsLIMITSetG/NG VoltageLimitsSetG/NG CurrentLimitsSetG/NG PowerLimitsL

Page 30 -   is OFF Key

Operation SLM-Series AC/DC Electronic Load 3-10 M540072-01 Rev B 3.3 NORMAL OPERATION 3.3.1 CC Mode: Two levels can be set to allow for quick switc

Page 31 - 3.2 SET-UP PROCEDURES

SLM-Series AC/DC Electronic Load Operation M540072-01 Rev B 3-11 3.3.3 CR Mode Two levels can be set to allow for quick switching between two curre

Page 32 - 3.2.2 Frequency Setting:

Operation SLM-Series AC/DC Electronic Load 3-12 M540072-01 Rev B MODEPress MODEButton until desiredmode is illuminatedCC, CRPRESLEVELFREQLEVELLOAD

Page 33 - 3.2.4 Crest Factor

SLM-Series AC/DC Electronic Load Operation M540072-01 Rev B 3-13 3.4 INITIAL SETTINGS OF SLM-SERIES AC/DC ELECTRONIC LOAD The initial setting par

Page 35 - DESCRIPTION DISPLAY CONTROL

Operation SLM-Series AC/DC Electronic Load 3-14 M540072-01 Rev B Description SLM-60-20-300 SLM-150-8-300 SLM-300-4-300 SLM-500-1-300 MODE CC CC

Page 36 - 3.3 NORMAL OPERATION

SLM-Series AC/DC Electronic Load Operation M540072-01 Rev B 3-15 SLM-300-4-300 Range 1 Range 2 FULL SCALE LOAD CURRENT 2 A 4 A CURRENT RANGE 4

Page 37 - 3.3.3 CR Mode

Operation SLM-Series AC/DC Electronic Load 3-16 M540072-01 Rev B 3.5.4 Over Temperature Protection (OTP) SLM-Series AC/DC Electronic load is provid

Page 38 - 3-12 M540072-01 Rev B

M540072-01 Rev B 4-1 SECTION 4 GPIB/RS-232 PROGRAMMING OPERATION 4.1 INTRODUCTION The rear panel of the SLM chassis is designed to connect with a PC

Page 39

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-2 M540072-01 Rev B The connections for the rear panel RS-232 interface are sho

Page 40

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-3 Description of GPIB Programming Command Syntax. { } :The

Page 41 - 3.5 PROTECTION FEATURES

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-4 M540072-01 Rev B Setting Preset Numeric Command Model Remark SLM DC SLD

Page 42

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-5 STATE Command Model RETURN SLM DC SLD SLM AC [STATe:]

Page 43 - GPIB/RS-232 PROGRAMMING

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-6 M540072-01 Rev B COMMAND NOTE RETURN [SYStem:] CHANnel{SP}{1|2|3|4}[A|B]{;|

Page 44 - Fig. 4.1b

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-7 REMARKS: 1. d : 0 - 9 2. GLOB : GLOBAL (ALL CHANNELS ACTI

Page 45

About Xantrex Xantrex Technology Inc. is a world-leading supplier of advanced power electronics and controls with products from small mobile units to

Page 46

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-8 M540072-01 Rev B CC:{SP}{NR2}{;|NL} This command is used to set the load cur

Page 47 - STATE Command Model RETURN

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-9 module will adjust the HIGH level to be equal to the LOW le

Page 48

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-10 M540072-01 Rev B POWER Level Purpose: The load power setting in Constant Po

Page 49 - 4.5.1 Setting Commands

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-11 LOAD OFF VOLTAGE Setting Purpose: The Load OFF voltage set

Page 50

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-12 M540072-01 Rev B CLEAR status register Purpose: CLEar the PROT and ERR stat

Page 51

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-13 SYNCHRONOUS ON/OFF Purpose: To set synchronous function ON

Page 52

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-14 M540072-01 Rev B Example: BANK 1, to set waveform bank 1. BANK 10, to set w

Page 53

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-15 3. If the value designated exceeds the specification of t

Page 54

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-16 M540072-01 Rev B VA Limit Purpose: To set the upper/lower limit value of th

Page 55

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-17 Description: The RISE load current slew rate of load level

Page 56

This page intentionally left blank. ii

Page 57

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-18 M540072-01 Rev B SHORT ON/OFF Purpose: Short the DC input of Electronic l

Page 58

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-19 voltage and current meter GO/NG check, so user can set the

Page 59

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-20 M540072-01 Rev B 4.5.2 Query Commands CHANNEL Purpose: The active channel

Page 60

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-21 VOLTAGE Level The Constant Voltage mode's load voltag

Page 61

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-22 M540072-01 Rev B LOAD OFF Voltage Purpose: LOAD OFF voltage level query com

Page 62 - 4.5.2 Query Commands

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-23 PROTECTION Status Register Purpose: OPP, OTP, OVP, and OC

Page 63

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-24 M540072-01 Rev B Example: Module SLM-60-60-300 MODE CC RANG 1 CURR:HIGH 2

Page 64

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-25 POWER METER Purpose: To read the value of Watt meter. Comm

Page 65

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-26 M540072-01 Rev B SYNCHRONOUS ON/OFF Purpose: To read the setting condition

Page 66

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-27 FREQUENCY Purpose: To read the set frequency of FREQ. Com

Page 67

Important Safety Instructions Before applying power to the system, verify that your product is configured properly for your particular application.

Page 68

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-28 M540072-01 Rev B NG Purpose: To read the set value of NG. Command Syntax:

Page 69

SLM-Series AC/DC Electronic Load GPIB/RS-232 Programming Operation M540072-01 Rev B 4-29 SHORT ON/OFF Purpose: SHORt ON or OFF status query comman

Page 70

GPIB/RS-232 Programming Operation SLM-Series AC/DC Electronic Load 4-30 M540072-01 Rev B This page intentionally left blank.

Page 71

M540072-01 Rev B A-1 APPENDIX A GPIB PROGRAMMING EXAMPLE Example Program in C /* Link this program with appropriate *cib*.obj. */ /* This applicat

Page 72 - 4-30 M540072-01 Rev B

Appendix A SLM-Series AC/DC Electronic Load A-2 M540072-01 Rev B ibrd(load,spec,20); if (spec[3] == '9') printf("\n SLM

Page 73 - GPIB PROGRAMMING EXAMPLE

SLM-Series AC/DC Electronic Load Appendix A M540072-01 Rev B A-3 Example Program in BASICA LOAD DECL.BAS using BASICA MERGE command. 100 REM You

Page 74

Appendix A SLM-Series AC/DC Electronic Load A-4 M540072-01 Rev B This page intentionally left blank.

Page 75 - Example Program in BASICA

M540072-01 Rev B B-1 APPENDIX B RS-232 PROGRAMMING EXAMPLE C Language Interface for DOS Handlers " pd_rs232.c " #include <dos.h> #i

Page 76 - A-4 M540072-01 Rev B

SLM-Series AC/DC Electronic Load Appendix B B-2 M540072-01 Rev B #define PE 4 /* parity error */ #def

Page 77 - RS-232 PROGRAMMING EXAMPLE

Appendix B SLM-Series AC/DC Electronic Load M540072-01 Rev B B-3 strupr(tempbuf); strupr(intbuf); cnt = 0; do { comerr

Page 78

This page intentionally left blank. iv

Page 79

SLM-Series AC/DC Electronic Load Appendix B B-4 M540072-01 Rev B queue[0] = '\0'; if (timeout == 1) { return(TIME_OUT

Page 80

Appendix B SLM-Series AC/DC Electronic Load M540072-01 Rev B B-5 break; } /* finally send the non-spacific */

Page 81

SLM-Series AC/DC Electronic Load Appendix B B-6 M540072-01 Rev B } strupr(measbuf); cnt = 0; do { measerr = strncmp(mea

Page 82

Appendix B SLM-Series AC/DC Electronic Load M540072-01 Rev B B-7 case '1': pd_wrt("chan 1",6);

Page 83

SLM-Series AC/DC Electronic Load Appendix B B-8 M540072-01 Rev B void pd_loc(void) { char ch; outportb(ACE_INT_ENB_REG, 0); /* disable a

Page 84 - C Example Program

Appendix B SLM-Series AC/DC Electronic Load M540072-01 Rev B B-9 /* Get the load actially sink current from the load */ delay(200); pd_rd(rdb

Page 85

SLM-Series AC/DC Electronic Load Appendix B B-10 M540072-01 Rev B This page intentionally left blank.

Page 86 - B-10 M540072-01 Rev B

M540072-01 Rev B C-1 APPENDIX C SLM-SERIES AC/DC LOAD GPIB/RS-232 OPERATING FLOW CHART REMSTARTCC MODE?CR MODE?NOSet"MODE CC"Set"MOD

Page 87 - OPERATING FLOW CHART

Product Family: SLM-Series AC/DC Electronic Load Warranty Period: One (1) Year WARRANTY TERMS Xantrex provides this written warranty covering the Pro

Commentaires sur ces manuels

Pas de commentaire