SL-SERIES SLM DC MODULAR/ SLH HIGH POWER DC ELECTRONIC LOADS Calibration Manual Elgar Electronics Corporation 92
Calibration Elgar Electronics Corporation 2.2.1 DVM, RANGE I OFFSET AND GAIN, ITEMS 1 AND 2 7. Set the DC Voltage Standard to the Input Voltage valu
Elgar Electronics Corporation Calibration 2.2.2 DVM, RANGE II OFFSET AND GAIN, ITEMS 3 AND 4 1. Press the © key to go to Item 3. 2. Set the DC Vol
Calibration Elgar Electronics Corporation 2.3 VSENSE CALIBRATION Make the connections for VSENSE calibration as shown in Figure 2-2. Figure 2-2. Co
Elgar Electronics Corporation Calibration 5. Set the DC Voltage Standard to the Input Voltage value shown in Table 2-6 for the model being calibrate
Calibration Elgar Electronics Corporation 5. Set the DC Voltage Standard to the Input Voltage value shown in Table 2-8 for the model being calibrate
Elgar Electronics Corporation Calibration 2.4 DAM AND CC MODE CALIBRATIONS The following three steps apply to both DAM and CC Mode Calibration, Item
Calibration Elgar Electronics Corporation 8. LOAD OFF. 9. Press the © key to go to Item 10. 10. PRES ON. 11. Set the load current as shown in Table
Elgar Electronics Corporation Calibration 2.4.2 DAM, RANGE II OFFSET AND GAIN, ITEMS 11 AND 12 1. LOAD OFF. 2. Press the © key to go to Item 11. 3.
Calibration Elgar Electronics Corporation 10. PRES ON. 11. Set the load current as shown in Table 2-12 for the model being calibrated. 12. PRES OFF.
Elgar Electronics Corporation Calibration 8. LOAD OFF. 9. Press the © key to go to Item 14. 10. PRES ON. 11. Set the load current as shown in Tabl
Calibration Elgar Electronics Corporation 2.4.4 CC MODE, LEVEL HIGH, RANGE I OFFSET AND GAIN, ITEMS 15 AND 16 1. LOAD OFF. 2. Press the © key to g
Elgar Electronics Corporation Calibration 16. PRES ON. 17. Set the load current as shown in Table 2-16 for the model being calibrated. 18. PRES OFF.
Calibration Elgar Electronics Corporation 8. LOAD OFF. 9. Press the © key to go to Item 18. 10. PRES ON. 11. Set the load current as shown in Tabl
Elgar Electronics Corporation Calibration 2.4.6 CC MODE, LEVEL HIGH, RANGE II OFFSET AND GAIN, ITEMS 19 AND 20 1. LOAD OFF. 2. Press the © key to
Calibration Elgar Electronics Corporation 10. PRES ON. 11. Set the load current as shown in Table 2-20 for the model being calibrated. 12. PRES OFF.
Elgar Electronics Corporation Calibration 2.5.1 CR MODE, LEVEL LOW, RANGE I OFFSET AND GAIN, ITEMS 21 AND 22 1. LOAD OFF. 2. Press the © key to go
Calibration Elgar Electronics Corporation 11. Set the power supply voltage and the DC Voltage Standard to the VSENSE Input value shown in Table 2-22
Elgar Electronics Corporation Calibration 2.5.2 CR MODE, LEVEL HIGH, RANGE I OFFSET AND GAIN ITEMS 23 AND 24 1. LOAD OFF. 2. Press the © key to go
Calibration Elgar Electronics Corporation 11. Set the power supply voltage and the DC Voltage Standard to the VSENSE Input value shown in Table 2-24
Elgar Electronics Corporation Calibration 2.5.3 CR MODE, LEVEL LOW, RANGE II OFFSET AND GAIN, ITEMS 25 AND 26 1. LOAD OFF. 2. Press the © key to
CONTENTS SECTION 1 OPERATION DESCRIPTION 1-1 1.1 NAVIGATION AND SETTING 1-2 SECTION 2 CALIBRATION 2-1 2.1 CALIBRATION SETUP 2-1 2.1.1 Equipment Re
Calibration Elgar Electronics Corporation 11. Set the power supply voltage and the DC Voltage Standard to the VSENSE Input value shown in Table 2-26
Elgar Electronics Corporation Calibration 2.5.4 CR MODE, LEVEL HIGH, RANGE II OFFSET AND GAIN, ITEMS 27 AND 28 1. LOAD OFF. 2. Press the © key to
Calibration Elgar Electronics Corporation 11. Set the power supply voltage and the DC Voltage Standard to the VSENSE Input value shown in Table 2-28
Elgar Electronics Corporation Calibration 2.6 CV MODE CALIBRATION Connect the DC Power Supply to the input terminal as illustrated in Figure 1-1. F
Calibration Elgar Electronics Corporation 2.6.1 CV MODE, LEVEL LOW, OFFSET AND GAIN, ITEMS 29 AND 30 For CV Mode Calibration, set the DC power suppl
Elgar Electronics Corporation Calibration 10. PRES ON. 11. Set voltage to the V Input value shown in Table 2-30 for the model being calibrated. 12.
Calibration Elgar Electronics Corporation 2.6.2 CV MODE, LEVEL HIGH, OFFSET AND GAIN, ITEMS 31 AND 32 Set the DC power supply current limit to 0.1A.
Elgar Electronics Corporation Calibration 10. PRES ON. 11. Set voltage to the V Input value shown in Table 2-32 for the model being calibrated. 12. P
Calibration Elgar Electronics Corporation 2.7.1 CP MODE, LEVEL LOW, OFFSET AND GAIN, ITEMS 33 AND 34 1. LOAD OFF. 2. Press the © key to go to Item
Elgar Electronics Corporation Calibration 9. Input Voltage to the VSENSE terminal per Table 2-34 for the model being calibrated. 10. PRES ON, settin
2.6 CV MODE CALIBRATION 2-25 2.6.1 CV Mode, Level Low, Offset and Gain, Items 29 and 30... 2-26 2.6.2 CV Mode, Level H
Calibration Elgar Electronics Corporation 2.7.2 CP MODE, LEVEL HIGH, OFFSET AND GAIN, ITEMS 35 AND 36 1. LOAD OFF 2. Press the © key to go to Item
Elgar Electronics Corporation Calibration 9. Input Voltage to the VSENSE terminal as Table 2-36. 10. PRES ON, setting Power as table 40, PRES OFF. 1
Calibration Elgar Electronics Corporation This page intentionally left blank. 2-34 SLM/SLH Calibration Manual
SECTION 3 CALIBRATION DATA RECORD Single Input DC Electronic Loads Calibration Report MODEL : SERIAL NO : DATE : Inspector : Item Descriptio
Calibration Data Record Elgar Electronics Corporation 3-2 SLM/SLH Calibration Manual
Table 2-18. CC Mode, Level Low, Range II Gain Adjustment Settings (Item 18)...2-14 Table 2-19. CC Mode, Level High, Rang
This page intentionally left blank. iv SLM/SLH Calibration Manual
SECTION 1 OPERATION DESCRIPTION This section provides a brief orientation for how to enter, navigate, adjust and store settings, and exit the calibra
Operation Description Elgar Electronics Corporation 1.1 NAVIGATION AND SETTING 1. Press MODE and LEVEL keys simultaneously to enter calibration mode
SECTION 2 CALIBRATION 2.1 CALIBRATION SETUP Prior to starting the calibration procedure, gather the necessary equipment (Section 2.1.1), then power u