SL-SERIES
SLM DC MODULAR/
SLH H
IGH POWER DC
E
LECTRONIC LOADS
Calibration Manual
Elgar Electronics Corporation
9250 Brown Deer Road
San Diego, CA 92121-2294
1-800-73ELGAR (1-800-733-5427)
Tel: (858) 450-0085
www.elgar.com
Fax: (858) 458-0267
©2006 by Elgar Electronics Corporation
This document contains information proprietary to Elgar Electronics Corporation. The information contained herein is not to be
duplicated or transferred in any manner without prior written permission from Elgar Electronics Corporation.
July 31, 2006 Document No. M540070-04 Rev A
Vue de la page 0
1 2 3 4 5 6 ... 43 44

Résumé du contenu

Page 1 - POWER DC

SL-SERIES SLM DC MODULAR/ SLH HIGH POWER DC ELECTRONIC LOADS Calibration Manual Elgar Electronics Corporation 92

Page 2

Calibration Elgar Electronics Corporation 2.2.1 DVM, RANGE I OFFSET AND GAIN, ITEMS 1 AND 2 7. Set the DC Voltage Standard to the Input Voltage valu

Page 3 - CONTENTS

Elgar Electronics Corporation Calibration 2.2.2 DVM, RANGE II OFFSET AND GAIN, ITEMS 3 AND 4 1. Press the © key to go to Item 3. 2. Set the DC Vol

Page 4

Calibration Elgar Electronics Corporation 2.3 VSENSE CALIBRATION Make the connections for VSENSE calibration as shown in Figure 2-2. Figure 2-2. Co

Page 5

Elgar Electronics Corporation Calibration 5. Set the DC Voltage Standard to the Input Voltage value shown in Table 2-6 for the model being calibrate

Page 6

Calibration Elgar Electronics Corporation 5. Set the DC Voltage Standard to the Input Voltage value shown in Table 2-8 for the model being calibrate

Page 7 - OPERATION DESCRIPTION

Elgar Electronics Corporation Calibration 2.4 DAM AND CC MODE CALIBRATIONS The following three steps apply to both DAM and CC Mode Calibration, Item

Page 8 - 1.1 NAVIGATION AND SETTING

Calibration Elgar Electronics Corporation 8. LOAD OFF. 9. Press the © key to go to Item 10. 10. PRES ON. 11. Set the load current as shown in Table

Page 9 - CALIBRATION

Elgar Electronics Corporation Calibration 2.4.2 DAM, RANGE II OFFSET AND GAIN, ITEMS 11 AND 12 1. LOAD OFF. 2. Press the © key to go to Item 11. 3.

Page 10

Calibration Elgar Electronics Corporation 10. PRES ON. 11. Set the load current as shown in Table 2-12 for the model being calibrated. 12. PRES OFF.

Page 11

Elgar Electronics Corporation Calibration 8. LOAD OFF. 9. Press the © key to go to Item 14. 10. PRES ON. 11. Set the load current as shown in Tabl

Page 13

Calibration Elgar Electronics Corporation 2.4.4 CC MODE, LEVEL HIGH, RANGE I OFFSET AND GAIN, ITEMS 15 AND 16 1. LOAD OFF. 2. Press the © key to g

Page 14

Elgar Electronics Corporation Calibration 16. PRES ON. 17. Set the load current as shown in Table 2-16 for the model being calibrated. 18. PRES OFF.

Page 15

Calibration Elgar Electronics Corporation 8. LOAD OFF. 9. Press the © key to go to Item 18. 10. PRES ON. 11. Set the load current as shown in Tabl

Page 16

Elgar Electronics Corporation Calibration 2.4.6 CC MODE, LEVEL HIGH, RANGE II OFFSET AND GAIN, ITEMS 19 AND 20 1. LOAD OFF. 2. Press the © key to

Page 17

Calibration Elgar Electronics Corporation 10. PRES ON. 11. Set the load current as shown in Table 2-20 for the model being calibrated. 12. PRES OFF.

Page 18

Elgar Electronics Corporation Calibration 2.5.1 CR MODE, LEVEL LOW, RANGE I OFFSET AND GAIN, ITEMS 21 AND 22 1. LOAD OFF. 2. Press the © key to go

Page 19

Calibration Elgar Electronics Corporation 11. Set the power supply voltage and the DC Voltage Standard to the VSENSE Input value shown in Table 2-22

Page 20

Elgar Electronics Corporation Calibration 2.5.2 CR MODE, LEVEL HIGH, RANGE I OFFSET AND GAIN ITEMS 23 AND 24 1. LOAD OFF. 2. Press the © key to go

Page 21

Calibration Elgar Electronics Corporation 11. Set the power supply voltage and the DC Voltage Standard to the VSENSE Input value shown in Table 2-24

Page 22

Elgar Electronics Corporation Calibration 2.5.3 CR MODE, LEVEL LOW, RANGE II OFFSET AND GAIN, ITEMS 25 AND 26 1. LOAD OFF. 2. Press the © key to

Page 23

CONTENTS SECTION 1 OPERATION DESCRIPTION 1-1 1.1 NAVIGATION AND SETTING 1-2 SECTION 2 CALIBRATION 2-1 2.1 CALIBRATION SETUP 2-1 2.1.1 Equipment Re

Page 24 - 2.5 CR MODE CALIBRATION

Calibration Elgar Electronics Corporation 11. Set the power supply voltage and the DC Voltage Standard to the VSENSE Input value shown in Table 2-26

Page 25

Elgar Electronics Corporation Calibration 2.5.4 CR MODE, LEVEL HIGH, RANGE II OFFSET AND GAIN, ITEMS 27 AND 28 1. LOAD OFF. 2. Press the © key to

Page 26

Calibration Elgar Electronics Corporation 11. Set the power supply voltage and the DC Voltage Standard to the VSENSE Input value shown in Table 2-28

Page 27

Elgar Electronics Corporation Calibration 2.6 CV MODE CALIBRATION Connect the DC Power Supply to the input terminal as illustrated in Figure 1-1. F

Page 28

Calibration Elgar Electronics Corporation 2.6.1 CV MODE, LEVEL LOW, OFFSET AND GAIN, ITEMS 29 AND 30 For CV Mode Calibration, set the DC power suppl

Page 29

Elgar Electronics Corporation Calibration 10. PRES ON. 11. Set voltage to the V Input value shown in Table 2-30 for the model being calibrated. 12.

Page 30

Calibration Elgar Electronics Corporation 2.6.2 CV MODE, LEVEL HIGH, OFFSET AND GAIN, ITEMS 31 AND 32 Set the DC power supply current limit to 0.1A.

Page 31

Elgar Electronics Corporation Calibration 10. PRES ON. 11. Set voltage to the V Input value shown in Table 2-32 for the model being calibrated. 12. P

Page 32

Calibration Elgar Electronics Corporation 2.7.1 CP MODE, LEVEL LOW, OFFSET AND GAIN, ITEMS 33 AND 34 1. LOAD OFF. 2. Press the © key to go to Item

Page 33 - 2.6 CV MODE CALIBRATION

Elgar Electronics Corporation Calibration 9. Input Voltage to the VSENSE terminal per Table 2-34 for the model being calibrated. 10. PRES ON, settin

Page 34

2.6 CV MODE CALIBRATION 2-25 2.6.1 CV Mode, Level Low, Offset and Gain, Items 29 and 30... 2-26 2.6.2 CV Mode, Level H

Page 35

Calibration Elgar Electronics Corporation 2.7.2 CP MODE, LEVEL HIGH, OFFSET AND GAIN, ITEMS 35 AND 36 1. LOAD OFF 2. Press the © key to go to Item

Page 36

Elgar Electronics Corporation Calibration 9. Input Voltage to the VSENSE terminal as Table 2-36. 10. PRES ON, setting Power as table 40, PRES OFF. 1

Page 37 - 2.7 CP MODE CALIBRATION

Calibration Elgar Electronics Corporation This page intentionally left blank. 2-34 SLM/SLH Calibration Manual

Page 38

SECTION 3 CALIBRATION DATA RECORD Single Input DC Electronic Loads Calibration Report MODEL : SERIAL NO : DATE : Inspector : Item Descriptio

Page 39

Calibration Data Record Elgar Electronics Corporation 3-2 SLM/SLH Calibration Manual

Page 40

Table 2-18. CC Mode, Level Low, Range II Gain Adjustment Settings (Item 18)...2-14 Table 2-19. CC Mode, Level High, Rang

Page 41 - 2.8 STORE CALIBRATION DATA

This page intentionally left blank. iv SLM/SLH Calibration Manual

Page 42

SECTION 1 OPERATION DESCRIPTION This section provides a brief orientation for how to enter, navigate, adjust and store settings, and exit the calibra

Page 43 - CALIBRATION DATA RECORD

Operation Description Elgar Electronics Corporation 1.1 NAVIGATION AND SETTING 1. Press MODE and LEVEL keys simultaneously to enter calibration mode

Page 44

SECTION 2 CALIBRATION 2.1 CALIBRATION SETUP Prior to starting the calibration procedure, gather the necessary equipment (Section 2.1.1), then power u